On the detection of non-methane total hydrocarbons by gas chromatography - Master's thesis - Dissertation
The gas chromatographic method of Junqi instrument for the determination of non-methane total hydrocarbons in fixed-source exhaust gas: 1 GC9890A gas chromatograph host (220V, 50HZ) *Imported HP high sensitivity FID detector * packed column inlet * packed column inlet *Double six-way injection valve * Five independent temperature control system, intelligent rear door *With data power-off protection function, fault self-diagnosis *Installing accessory kit 2 V2008 Chromatography Data Station Dual channel capture card and software (Double channel, gas-liquid universal) analysis software 3 Column Total hydrocarbon analysis packed column *Main features 5, can be connected to the counter-control workstation, real-time control, and can store 100 sets of experimental conditions and spectrum. 6, can be installed in the United States HP-5890 gas chromatograph micro-thermal conductivity detector, to achieve complete docking. * Heating furnace system temperature range: ambient temperature +5 ° C ~ 400 ° C third-order program temperature, heating rate 0 ~ 50 ° C / min; increase The amount is 0.1 ° C / min. Temperature control accuracy ± 0.05 ° C, the user can re-correct the furnace temperature, and set the maximum temperature. The intelligent rear door opening system has stepless variable inlet and outlet air volume, which shortens the system stable balance time after rising/lowering. The time taken for the oven to drop from 300 ° C to 50 ° C is < 6 min. *Injection system 1. Packed column sampler: suitable for various caliber packed columns and large diameter capillary columns 2. Capillary injector: optional split/distributed system controlled by programmed valve 3, optional with the company's headspace sampler and thermal analyzer. *Detector system Easy to disassemble and install, easy to clean or replace nozzles, easy to operate; input signal can be logarithmically amplified, reducing interference, high sensitivity, good linearity and wide measuring range. 2. Thermal Conductivity Detector (TCD): Sensitivity: Type A: ≥4000mv•ml/mg (n-hexadecane) Baseline noise: ≤30μv (helium gas with a carrier gas of 99.999) Our Lithium Battery includes 5G Base Station Backup Power System,like 48V 100Ah/150Ah/200Ah Lithium Battery. 3.2V Prismatic cells,like 3.2V 50Ah/105Ah/202Ah Lithium Battery. And Lithium Ion Pouch Cells, including 3.2V 12-30Ah. Lifepo4 Battery,Lifepo4 Lithium Ion Battery,Lifepo4 48V 100Ah Lithium Ion Battery,Lithium Ion Battery For Solar 100Ah Jiangsu Zhitai New Energy Technology Co.,Ltd , https://www.jszhitaienergy.com
1. The main machine comes with two-way valve control system, which is convenient for online automatic injection.
2. The inter-column compensation circuit electronically compensates for drift in the baseline during program temperature operation, reducing the complexity of the second column, detector, and auxiliary flow system.
3. The signal of the detector, the value of the heater, the furnace temperature, the flow sensor reading or the stored column compensation baseline signal can all be assigned to an analog output channel.
4, self-machine detection and fault diagnosis, power-off protection stored experimental data, stopwatch and running timer, keyboard lock function.
7, can install two flow sensing channels, two single column compensation channels.
1. Hydrogen flame ion detector (FID):
Sensitivity: Type A Mt ≤ 5 × 10-11g / s (n-hexadecane)
Baseline noise: ≤6×10-12A/H
Linear range: ≥105
Stabilization time: <0.5h
Stabilization time: A type <1.5h gas barrier protection